Main technical performances
1. Category: High vacuum / high flow
2. Power Supply: AC(100~240)V ,50Hz/60Hz or DC 12V 5A
DC power supply The DC power supply and Mains circuit meets the
requirements of the 2MOOP
3. Input power: 150VA
4. *Maximum vacuum: (85±5)kPa
5. Vacuum regulating range: 20kPa~Maximum vacuum
6. *Free air flow: (27±4)l/min **
7.Collection container: 1000ml/pc,1pc
8. Noise: ≤65dB(A)
9. Weight: Approx 4kg
10. Dimension(approx): 480 mm(L)×189 mm(W)×285 mm(H)
11. Service life: 5 years (except for fragile and consumable parts)
12. Battery operating time: ≥60min (Battery operating time during continu
ous operation with fully recharged battery /new battery setting at limit nega
tive pressure.
13. Mode of operation: non-continuous operation 30 minutes on, 30minutes
off.
14. Electric shock protection category: Class II external power supply
equipmentIngress of liquid protection category: IP21.
15. Electric shock protection degree: BF type applied parts Applied part
(Suction catheter)
16. The suction unit is not suitable for use in the place with inflammable
&explosive gas.
17. Air filter service life: two weeks, Fuse and Power cord service life:
5 years.
18. Detachable components of Portable Phlegm Suction Unit shelf-life:
See inner packing bag.
19. Fuse: F1.6AL 250V, Ф5×20(SUPLLY MAINS)
*Test conditions: 760mmHg (at one atmosphere pressure)
** Measuring point at the device inlet, cold machine status
Normal operating condition
Ambient temperature: +5℃~+35℃
Relative humidity: 30%~80%
Atmosphere pressure: 86kPa~106kpa
NOTE: When storage temperature is below 5℃ or above 35℃,please keep the equipment in normal working condition for at least 4 hours before using.
NOTE: Not intended to be used in an MRI environment.