General Features
Portable Suction Unit is intended for negative pressure supply,and is used in combination with suction tubing and suction catheters to remove fluids and debris from the surgical field during surgical procedures. It is suitable for patients who need to aspirate purulent and viscous fluids.
Structure&working principle
1.Oil free lubrication pump to keep the environment from being polluted by the oil mist.
2.Low noise.
3.Round negative pressure meter, and plastic cover.
4.No any positive pressure to be generated during running, to ensure
reliable and safe operation.
5.Negative pressure regulating system can be adjusted steplessly. Small in size, light in weight and portable.
Main technical performances
1. Category: High vacuum / high flow
2. Power Supply: AC(100~240)V ,50Hz/60Hz or DC 12V 5A
DC power supply The DC power supply and Mains circuit meets the
requirements of the 2MOOP
3. Input power: 150VA
4. *Maximum vacuum: (85±5)kPa
5. Vacuum regulating range: 20kPa~Maximum vacuum
6. *Free air flow: (27±4)l/min **
7.Collection container: 1000ml/pc,1pc
8. Noise: ≤65dB(A)
9. Weight: Approx 4kg
10. Dimension(approx): 480 mm(L)×189 mm(W)×285 mm(H)
11. Service life: 5 years (except for fragile and consumable parts)
12. Battery operating time: ≥60min (Battery operating time during continu
ous operation with fully recharged battery /new battery setting at limit nega
tive pressure.
13. Mode of operation: non-continuous operation 30 minutes on, 30minutes
off.
14. Electric shock protection category: Class II external power supply
equipmentIngress of liquid protection category: IP21.
15. Electric shock protection degree: BF type applied parts Applied part
(Suction catheter)
16. The suction unit is not suitable for use in the place with inflammable
&explosive gas.
17. Air filter service life: two weeks, Fuse and Power cord service life:
5 years.
18. Detachable components of Portable Phlegm Suction Unit shelf-life:
See inner packing bag.
19. Fuse: F1.6AL 250V, Ф5×20(SUPLLY MAINS)
*Test conditions: 760mmHg (at one atmosphere pressure)
** Measuring point at the device inlet, cold machine status
Normal operating condition
Ambient temperature: +5℃~+35℃
Relative humidity: 30%~80%
Atmosphere pressure: 86kPa~106kpa
NOTE: When storage temperature is below 5℃ or above 35℃,please keep the equipment in normal working condition for at least 4 hours before using.
NOTE: Not intended to be used in an MRI environment.